Two integrating-sphere reflectance reports can give different values for the same surface without either instrument being wrong. The methods may collect different parts of the reflected light. Before setting an acceptance limit for a coated or moulded optical surface, say whether the specular return is included and define the setup that goes with the number.

Commercial integrating sphere with a visible measurement port
Commercial integrating sphere example, not a SilverOptics instrument or test setup. Photograph by Electro Optical Industries, Wikimedia Commons, CC BY-SA 3.0.

Name the quantity before setting the limit

State whether the test should include the mirror-like return or exclude it. Also specify illumination wavelength or band and incidence geometry. A glossy surface can send much of its return into a narrow direction; a textured surface distributes more of it over other angles. This is why a diffuse-only result cannot be compared directly with one that includes the specular component.

Put port state and sample placement in the report

Ask the lab to identify the measurement configuration, including how the specular return is collected or excluded, the sample orientation and the reference used. For a curved or undersized sample, check with the lab that the illuminated area lies on the intended surface and that the holder or gap does not introduce a different light path. If two labs report different values, compare this setup before changing the material or coating specification.

La NIST summary of directional-hemispherical and diffuse reflectance experiments explains why the size of the specular-exclusion opening can influence measured diffuse values. The example is infrared; it does not set an acceptance range for a visible-light product.

Ask for comparable data across lots

Define the relevant drawing revision, surface, measurement locations, number of parts and reporting format. Save a representative report with its configuration at initial approval so changes in later lots can be separated from changes in method. If reflectance is part of an inquiry, use the optical coating requirements page to identify the coating and operating conditions, and confirm the specific measurement scope with the lab before release.